• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 5-1. Operator's Preventive Maintenance Checks and Services
Table 5-1 Operator's Preventive Maintenance Checks and Services - Continued

Maintenance Manual For Tester, Density And Moisture (Soil And Asphalt) Nuclear Method Page Navigation
  44    45    46    47    48  49  50    51    52    53    54  
img
. . . .
TM 5-6635-386-12&P
Table 5-1 Operator's Preventive Maintenance Checks and Services
Item
Item to
Not Fully Mission
No.
Interval
Check/ Service
Procedure
Capable If:
2
During
End Item
Diagnostic Inspection, check the
following:
a. Does the display work on all
Display not working.
channels?
b. Does the tester beep during
Tester doesn't beep
counting?
during counting.
c. Does the tester count in TEST '
Tester doesn't
MODE? What does it indicate
count/read-out
upon completion? Section 1,
incorrect.
Para 1-4.
d. Does it count in STANDARD
Tester doesn't
MODE? Is this a normal standard
count/read-out
count? What is the spread of the
incorrect.
standard count? Refer to Section1,
Para 1-7 and 1-8.
e. Does it count in the field count
Tester not counting
conditions? In all time sequences?
in all time sequences
3
After
Batteries
Check the batteries display light
"L" or nothing in
window, if "L" is displayed recharge
display light window.
batteries overnight using charger.
Charge NICAD batteries with AC
charger or DC leads. (Refer to
illustration, Page 5-7).
5-6


Privacy Statement
Press Release
Contact